QuantumCore Inc. · NbTiN device run · NEMO cleanroom
wafers in fab
27/33
blocks in fab
6/6
Pick a step to read the tooling, chemicals and procedure straight from the SOP.·4-inch Si → NbTiN → patterning → 3×3 blocks → measurement
data source
Live Google Sheets
last successful sync
Oct 13, 2026, 01:10 AM
reporting week
Oct 13, 2026
next refresh
Oct 20, 2026
records loaded
33 wafers · 6 blocks · 24 cooldowns
measurement log
Cryogenic device measurements
One entry per cooldown, transcribed from the measurement workbook (Oct 2025 – Aug 2026).
sessions
24
with gain
5
dead chips
4
best gain
20 dB
fridges
7
24 of 24 sessions
Partial / features onlySep 28, 2026- 17nm thick, 120mm long TWPA
-17nm thick, 50 ohm straight line
Both films are made with 200 nm thick a-Si but designed for 100 nmWH-B3-20G3-12
fridge
ARS
sample holder
SH5x14_2
SH5x14_3
station
—
logbook
20260930
thickness
17 nm
line length
120 mm
stopband
—
gain
—
runsARS_firstcooldown_WH_B3_20G3
Awesome results, stop band near 14.5 GHz. Clear gain seen near 25-30 dB at certain frequencies. Pumping of the cryostat compressor was noticed to move the TWPA transmission up and down with an unknown cause. This was not noticed with the through line chip
blocks
Partial / features onlySep 21, 202617nm-thick and 100mm-long(WH-B3-20G3-11) and 17nm-thick and 120mm-long(WH-B3-20G3-12). Both chips have 200nm-thick a-Si and designed for nm-thick film.WH-B3-20G3-11WH-B3-20G3-12
fridge
ZPC1
sample holder
SH5x14_1(WH-B3-20G3-11) SH5x14_2(WH-B3-20G3-12)
station
—
logbook
WH_B3_20G3_11and12_zpc1_20260921
thickness
17 nm
line length
100 / 120 mm
stopband
14.5 GHz
gain
—
runsWH_B3_20G3_11and12_zpc1_1_20260921
Both devices have have smooth transmission with a stopband ~14.5GHz. For both devices We see a clear gain profile by subtacting transmission w/o pump however the 0 gain level is shifted by ~-20dB due to attenuation of the signal when the pump is on.
blocks
Partial / features onlySep 15, 202617nm, 100mm, from old batch deposited 02112026W0D-B1-20G2U4
fridge
He Dewar
sample holder
SH3
station
—
logbook
20260915
thickness
17 nm
line length
100 mm
stopband
—
gain
—
runsRAC_dewar_W0D_B2_20GU4_17nm_100mm
Stop-band location replicates ZPC1 measurement. Sensible pump sweep results with gain ~ 5 dB. TDR suggests high impedance and loss
blocks
Partial / features onlySep 4, 2026W0D-B2-20G0U4-11 is 17nm-thick and 100mm-long device from the old batch and old design with 10Ohm stripline. WG-B2-8G1-31 is 8nm-thick and 20mm-long device from the new batch but old design.WG-B2-8G1-31W0D-B2-20G0U4-11
WG-B2-8G1-31 has bad transmission with high losses. W0D-B2-20G0U4-11 has a good transmission with a stopband ~10.2GHz. We see a clear gain profile by subtacting transmission w/o pump however the 0 gain level is shifted ~20dB down.
blocks
Partial / features onlySep 1, 202617nm, 50mm, from old batch deposited 02112026W0D-B1-20G2U4
fridge
RAC
sample holder
RAC
station
—
logbook
20260901.pdf
thickness
17 nm
line length
50 mm
stopband
—
gain
—
runsRAC_17nm_50mm_W0D-B1-20G2
Remeasure of working device. Results were replicated. TDR was done and we saw very high impedance and loss
blocks
Partial / features onlyAug 27, 2026WG-B2-8G1-22 (8nm-thick and 100mm-long, 21 supercells) and WF-2D-20G2-? (20nm-thick and 100mm-long)WG-B2-8G1-22WF-2D-20G2-??
fridge
ZPC1
sample holder
SH3 (WG-B2-8G1-22), SH4 (WF-2D-20G2-??)
station
Measurement laptop
logbook
WG_B2_8G1_and_WF_2D_20G2_zpc1_20260827
thickness
8 nm
line length
100 mm
stopband
4 GHz
gain
—
runsWG_B2_8G1_and_WF_2D_20G2_zpc1_1_20260827
Both devices have bad transmission. The transmission in WG-B2-8G1-22 has some stopband-looking features and high losses from 4GHz to 6GHz similar to WG-B2-8G1-21. The transmission in WF-2D-20G2-?? is lossy without pronounced features. For both devices we don't see a dependence of gain on pump frequency.
blocks
Dead chipAug 25, 20268nm, 100mm, 19 cells/scWG-B2-8G1
fridge
RAC
sample holder
RAC
station
—
logbook
20260825.pdf
thickness
8 nm
line length
100 mm
stopband
—
gain
—
runsRAC_8nm_100mm_WG_B2_8G1
Dead chip, calibrated TDR results show a short very close to the beginning of the line, although the results are slightly asymmetrical for Port 1 and Port 2
blocks
Dead chipAug 11, 20268nm, 50mm, 19 cells/scWG-B2-8G1
fridge
RAC
sample holder
RAC
station
—
logbook
20260811.pdf
thickness
8 nm
line length
50 mm
stopband
—
gain
—
runsRAC_8nm_50mm_19
Dead chip, uncalibrated TDR results suggest a short very close to the beginning of the line
blocks
Gain observedAug 9, 2026WG-B2-8G1-21 (8nm-thick and 50mm-long) and WF-2D-20G2 (20nm-thick and 50mm-long)WG-B2-8G1-21WF-2D-20G2-??
WF-2D-20G2-?? device transmisson doesn't show a stopband but we see a 20dB gain for signal band with a stopband-looking feature ~7.8GHz. We see a jump in transmission power level for a weak pump tone signal. WG-B2-8G1-21 device has bad transmission with high losses from 4GHz to 6GHz. Both devices demonstrate 4WM processes.
blocks
Partial / features onlyJul 24, 202617nm, 50mm, from old batch deposited 02112026W0D-B1-20G1
fridge
RAC
sample holder
RAC
station
—
logbook
20260724.pdf
thickness
17 nm
line length
50 mm
stopband
10.3 GHz
gain
—
runsRAC_17nm_50mm_old
Good results, stopband near 10.3 GHz
blocks
Gain observedJul 20, 202620-nm thick, 50mm-long devices measuered on 6/5/2026W0C-B2-20G1U1-32
W0C-B2-20G1U1-32 is connected to a new setup that filters out idler and pump band. We observe a 20dB gain. We observe a jump in power level of the device's transmission after applying a weak pump tone. Also, we see a similar jump for the temperature dependence between 445mK and 1.57K.
blocks
Dead chipJul 15, 202617nm, 150mmW0A-B1-20G2
fridge
RAC
sample holder
RAC
station
—
logbook
20260715.pdf
thickness
17 nm
line length
150 mm
stopband
—
gain
—
runsN/A
Dead chip
blocks
Gain observedJun 24, 2026One is 20-nm thick, 50mm-long devices measuered on 4/17/2026 and 6/5/2026. Second is 17-nm thick, 175mm-long (W0A-B1-20G1U2-11)W0C-B2-20G1U1-21W0A-B1-20G1U2-11
W0A-B1-20G1U2-11 device's transmission shows high losses indicating either short or open circuit. Old 20nm-thick device is connected to a new setup that filter out idler and pump bandwidth. We observe a jump in power level of the device's transmission after applying a weak pump tone (similar to what we see in 6/5/2026). Thus, the gain is either <10dB if we subtract transmission after the jump or 25dB if we subtract unpumped transmission.
blocks
Dead chipJun 13, 2026One is 20-nm thick, 50mm-long devices measuered on 4/17/2026 and 6/5/2026. Second is 17-nm thick, 50mm-long (W0A-B1-20G1U2-31)W0C-B2-20G1U1-21W0A-B1-20G1U2-31
fridge
ZPC2
sample holder
SH3, Alex's SH
station
Measurement laptop
logbook
QC_17nm_50mm_ZPC2_20260613
thickness
20 nm
line length
50 mm
stopband
—
gain
—
runs50mm_17nm_zpc2_1_20260613
W0A-B1-20G1U2-31 device has bad transmission indicating either short or open circuit. Old 20nm-thick device confirmed results observed in 6/5/2026
blocks
Gain observedJun 10, 202620-nm thick, 100mm-long devices from the same butch as from 4/17/2026W0C-B2-20G1U1-31
fridge
ZPC2
sample holder
SH3
station
Measurement laptop
logbook
QC_20nm_100mm_ZPC2_20260610
thickness
20 nm
line length
100 mm
stopband
10.5 GHz
gain
15 dB
runs100mm_20nm_1_20260610 · 100mm_20nm_3_20260611
Device is from the same batch as the one where we observed 15dB of gain. Transmission shows high losses with a stopband-looking feature ~10.5GHz. We also see not frequency dependent gain for random frequency ranges.
blocks
Gain observedJun 5, 2026Two 20-nm thick, 50mm-long devices. One was measured on 4/17/2026 and the other is from the same batch.W0C-B2-20G1U1-21W0C-B2-20G1U1-32
fridge
ZPC2
sample holder
SH1, Alex's SH
station
Measurement laptop
logbook
QC_20nm_50mm_ZPC2_20260605
thickness
20 nm
line length
50 mm
stopband
10.5 GHz
gain
15 dB
runs20260605
Device in Alex's SH was measured on 4/17/2026. Device in SH1 was measured in RAC's probe. For both devices we see a stopband ~10.5GHz and ~10dB of gain in the range 4to8GHz and ~15dB of gain in the range 6 to 8 GHz. Device in Alex's SH has a weired transmission with jumps in power until we apply a weak random pump tone.
blocks
Partial / features onlyMay 1, 202620-nm thick, 50mm-, 100mm-, 150mm- and 175mm-long devices. New design without long 10Ohm stripline.W0C-B3-20G1U2-??W0C-B3-20G1U2-??W0C-B3-20G1U2-??W0C-B3-20G1U2-??W0C-B3-20G1U2-??
Device from the 3/22/2026 experiment is connected to a new setup that filters out idler and pump band. We observed a small gain but no improvement comparing with a simple setup.
blocks
Partial / features onlyMar 22, 202617-nm thick, 50mm-long deviceW0D-B1-20G0U3-11
Device has a stopband ~8.6GHz and shows a small gain and 4WM frequency conversion.
blocks
Partial / features onlyFeb 23, 2026Two 8nm-thick, 20mm-long device is exposed with eBeam. One is measured on 12/17/2025 and the other is from the same batch.?-?-?-?
fridge
Tremblant
sample holder
Vadiraj's SH (old 20mm-long device), SH1 (new 20mm-long device)
20mm-long device from 12/17/2025 is connected to a new setup that filters out signal band. Observed a small gain and frequency connection. Transmission of the new 20mm-long device has a stopband 6GHz and ripples. Device shows a small gain and 4WM frequency conversion.
Partial / features onlyJan 7, 2026Two 8nm-thick, 10mm-long device is exposed with MLA with narrow and wide stubs.?-?-?-?
fridge
Tremblant
sample holder
SH1 and Alex's SH.
station
Whistler PC
logbook
QC_10_MLA_20260107
thickness
8 nm
line length
10 mm
stopband
—
gain
—
runsQC10_3_1_20260107
We did two cooldowns. Nothing interesting except "narrow" device had some stop-band looking featuer ~11.5GHz. Both devices had big ripples.
Partial / features onlyDec 17, 20258nm-thick, 20mm- and 10mm-long. 10mm-long device is exposed with MLA and had wrong connection inside the fridge.?-?-?-?
10mm-long device doesn't have a stopband but we see some pump frequency dependence and gain. Also, we see 4WM conversion. 100mm-long device doesn't show pump frequency dependence and a stopband.